2008 International Conference on BioMedical Engineering and Informatics Development of Precise-temperature-controlled Cooling Apparatus for Medical Application by Using Peltier Effect May 27-May 30 ISBN: 978-0-7695-3118-2
DOI Bookmark: http://doi.ieeecomputersociety.org/10.1109/BMEI.2008.239
This paper describes precise temperature controlled cooling apparatuses for medical application. A Peltier cryoprobe for dermis treatment and flexible cryoprobe for cryosurgery for deep region of human body were developed by utilizing the Peltier effect. The evaluation experiment of the Peltier cryoprobe was conducted. The proposed apparatus has enough performance to necrotize target diseased cells precisely. The proposed flexible cryoprobe is composed of a flexible resin tube enough to treat the deeper parts of the human body with a catheter. A Peltier element is attached at the top of the probe for active heat-transfer control of the cooling tip. The cooling experiments are carried out and the performance of the cryoprobe is evaluated. The achieved minimum temperature at the tip is -43 ˚C.
Citation:
Shigenao Maruyama, Atsuki Komiya, Hiroki Takeda, Setsuya Aiba, "Development of Precise-temperature-controlled Cooling Apparatus for Medical Application by Using Peltier Effect," bmei, vol. 2, pp.610-614, 2008 International Conference on BioMedical Engineering and Informatics, 2008 Usage of this product signifies your acceptance of the Terms of Use. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||