loading...
 This Article 
   
 Share 
   
 Bibliographic References 
   
 Add to: 
 
Digg
Furl
Spurl
Blink
Simpy
Google
Del.icio.us
Y!MyWeb
 
 Search 
   
2008 International Conference on BioMedical Engineering and Informatics
Development of Precise-temperature-controlled Cooling Apparatus for Medical Application by Using Peltier Effect
May 27-May 30
ISBN: 978-0-7695-3118-2
This paper describes precise temperature controlled cooling apparatuses for medical application. A Peltier cryoprobe for dermis treatment and flexible cryoprobe for cryosurgery for deep region of human body were developed by utilizing the Peltier effect. The evaluation experiment of the Peltier cryoprobe was conducted. The proposed apparatus has enough performance to necrotize target diseased cells precisely. The proposed flexible cryoprobe is composed of a flexible resin tube enough to treat the deeper parts of the human body with a catheter. A Peltier element is attached at the top of the probe for active heat-transfer control of the cooling tip. The cooling experiments are carried out and the performance of the cryoprobe is evaluated. The achieved minimum temperature at the tip is -43 ˚C.
Citation:
Shigenao Maruyama, Atsuki Komiya, Hiroki Takeda, Setsuya Aiba, "Development of Precise-temperature-controlled Cooling Apparatus for Medical Application by Using Peltier Effect," bmei, vol. 2, pp.610-614, 2008 International Conference on BioMedical Engineering and Informatics, 2008